CME Standard Upright Pharmacy Refrigerator Certified to NSF/ANSI 456 12 cu. ft. capacity

Details

CME Standard Upright Pharmacy Refrigerator Certified to NSF/ANSI 456 12 cu. ft. capacity

These cutting-edge pharmacy refrigerators are certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. With this certification, units protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery. Our Standard line provides multi-functional features in a cost-effective design.

These glass door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, and probe access ports with included probes. Units run on natural, hydrocarbon refrigerant for environmental health and energy efficiency.

Write a review
In stock
SKU
MPN
-
$5,961.71
Specifications
DiscontinuedNo
Additional Specifications:
DIMENSIONS 25" (W) x 29" (D) x 65 3/4" (H)
DOOR SWING 23 1/4"
OPERATIONAL ENVIRONMENT INDOOR USE ONLY, +18°C TO +26°C (+65°F TO +78°F), <70% RH
STORAGE CAPACITY 12 CU. FT. GROSS VOLUME
DOOR ONE SWING, SELF-CLOSING, NON-REVERSIBLE, MAGNETIC SEALED GASKET, KEYED LOCK
 
SHELVES FOUR SHELVES (THREE ADJUSTABLE/ONE FIXED) WITH GUARD RAIL ON BACK
MOUNTING 3 1/2" SWIVEL CASTERS(TWO LOCKING)
INTERIOR LIGHTING SHIELDED, SWITCHED LED LIGHTING, FULL COVERAGE, BALANCED SPECTRUM
AIRFLOW MANAGEMENT FORCED AIR TECHNOLOGY, PATENT PENDING
EXTERNAL PROBE ACCESS REAR WALL PORT (3/4") DIA.
INSULATION CABINET IS FOAMED-IN-PLACE WITH EPA COMPLIANT HIGH DENSITY URETHANE FOAM
EXTERIOR MATERIALS WHITE POWDER COATED STEEL
ACCESS CONTROL PYXIS®, OMNICELL® AND ACUDOSE RX® COMPATIBLE
GENERAL WARRANTY ONE (1) YEAR PARTS AND LABOR WARRANTY, EXCLUDING DISPLAY PROBE CALIBRATION
COMPRESSOR WARRANTY FIVE (5) YEARS COMPRESSOR WARRANTY
PRODUCT WEIGHT 231 LBS.                
SHIPPING WEIGHT 256 LBS.                
RATED AMPERAGE 3 AMPS                
POWER PLUG/POWER CORD NEMA 5-15 PLUG, 8 TO 10 FT TYPICAL, CONFORMS TO UL471 REQUIREMENTS, VACCINE STORAGE POWER CORD WARNING LABEL
FACILITY ELECTRICAL REQUIREMENT 110-120V AC: 15 A (MINIMUM)
AGENCY LISTING AND CERTIFICATION CERTIFIED WITH THE TEMPERATURE PERFORMANCE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE FOR ALL TESTING SCENARIOS. UL, C-UL, ETL, C-ETL LISTED AND CERTIFIED TO UL471 STANDARD, HYDROCARBON REFRIGERANT SAFETY.
INCLUDED ACCESSORIES TEMPERATURE MONITOR DEVICE (TMD) COMPLIES WITH THE CURRENT CDC GUIDELINES, WITH 3 YEARS CERTIFICATION OF CALIBRATION, “BUFFERED” PROBE IN THE PRODUCT SIMULATED SOLUTION, MIN/MAX MEMORY, FIELD INSTALLABLE, AND VISUAL & AUDIBLE TEMP ALARM

PHARMACY REFRIGERATOR/FREEZER TOOLKIT AND TEMPERATURE LOGS